PEF:
FEV1:
FEV6:
FEV1/FEV6:
Accuracy:
Sensor:
Memory :
Colour zones:
Reference values:
Quality factor:
Sounds:
Communication:
External settings:
Battery:
Battery life:
Dimensions:
Weight:
Accessories:
Resistance:
Operating temp.:
Storage temp.:
Humidity:
Barometric:
Performance:
Safety:
Regulatory:
Contracts:
USA Patent:
Technical Specifications
Range 15 - 999 LPM, (1 LPM resolution)
Range 0.15 - 9.99 litre, (0.01 litre resolution)
Range 0.15 - 9.99 litre, (0.01 litre resolution)
Range 0.01 - 0.99, (0.01 resolution)
PEF: 5% or +/- 20 LPM, (whichever is greater)
FEV1: +/- 3.5% or 0.1 litre, (whichever is greater)
FEV6: 4% or 0.1 litre (whichever is greater)
Pressure/flow sensor technology (patented)
96 test scores containing: PEF, FEV1, FEV6,
FEV1/FEV6, Colour Zone, Q-Factor
3 colour zones (Green, Yellow, Red)
Set by user for PEF (for PiKo-1) and FEV/FEV6
(for PiKo-6); Default settings: PiKo-1 = PEF 500 LPM,
PiKo-6 = FEV1/FEV6 0 (disabled)
Warning and indicator for cough or abnormal blow
4 patterns for different indications and warnings
Bi-Directional IR port
Possible settings using the optional PiKoNET
Professional software:
- Colour zone limits (in 10% increments)
- Use of FEV1 for PiKo-1 and FEV1 or FEV6 for PiKo-6
as an alternate reference type
2x type 357 silver oxide button cells (or equivalent)
PiKo-1: approx. 1 year; PiKo-6: approx. 6 months
(based on average of 6 blows/day)
75 X 35 X 20 mm, (3 X 1.3 X 0.8 inches)
35 grams (1.25 oz)
Plastic Mouthpieces - part number:393535
< 2.5 cm H
10 to 38
(-20) to 60
0 - 100% relative humidity
550 to 780 mmHg
ATS1994 (monitoring), AS/NZS-4237: 1994,
EN13826: 2003
EN60601-1, EN60601-1-1, EN60601-1-2, EN13826:2003,
Type BF patient applied part, and IPX4
PiKo-6 FDA 510(k); PiKo-1 FDA 510 (k) for OTC;
CE (0086), Class I with measurement function
GSA V797P-4063a; UK NHS Drug Tariff prescribable on
FP10
US # 6,447,459 B1
O/l/S @ 14 LPS or lower
2
C (50 to 100
F)
o
o
C (-4) to 140
F)
o
o
- 9 -