MSA ALTAIR Pro Manual De Operación página 8

Ocultar thumbs Ver también para ALTAIR Pro:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Do not use pressurized air to clean the sensor or
alarm holes.
All instrument readings and information must be
interpreted by someone trained and qualified in
interpreting instrument readings in relation to the
specific environment, industrial practice and exposure
limitations.
ALTAIR Pro Detector's gas sensors may have a cross-
sensitivity response to common interference gases
other than the gas it is designed to detect. In some
cases, this cross-sensitivity may be positive,
producing a reading higher than actual. In other cases,
it may be negative, giving lower readings or even
negative. Contact MSA for further details if cross-
sensitivity may be an issue.
Special requirements for ALTAIR PRO O 2 -R Remote
Oxygen Detector:
Operate this instrument only in ambient
temperatures ranging from 0ºC to 40ºC
(32ºF to 104ºF).
If the instrument is taken into an environment
where the ambient temperature is 15ºC (59ºF) or
more, higher or lower, than its previous location,
do not use the instrument for a period of
40 minutes to allow the sensor to adjust to the
new conditions. After this 40-minute period,
perform a Fresh Air Set-up (FAS) to re-zero the
sensor. Perform the FAS in fresh air only.
Quick changes in barometric pressure result in
an offset in the sensor reading for a few
minutes. If this occurs, do not use the
instrument for a period of five minutes; then,
perform a Fresh Air Set-up (FAS) to re-zero the
sensor. Perform the FAS in fresh air only.
Conditions for Safe Use:
When it is necessary to replace the battery, use
only a Lithium battery listed in TABLE 6-2.
Use only the specified MSA cable connecting the
remote Oxygen sensor.
The maximum length of the remote sensor
connecting cable must not exceed 3.3 meters
(10 feet).
INCORRECT USE OR FAILURE TO FOLLOW THIS WARNING
CAN RESULT IN SERIOUS PERSONAL INJURY OR DEATH.
1-2

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair pro

Tabla de contenido