E01
• Service icon is lit
• Continuous acoustic alarm
• E01 flashes in the actual pressure
display.
E02
• Service icon is lit
• Continuous acoustic alarm
• E02 flashes in the actual pressure
display
E50
• Service icon is lit
• Acoustic alarm, 1x
• E50 flashes in the actual pressure
display
E51
• Service icon is lit
• Acoustic alarm, 2x
• E51 flashes in the actual pressure
display
E59
• Acoustic alarm, 10x
• E59 flashes in the actual pressure
display
• Safe status, only M key is active
No display in the display unit.
• Pressure measuring failure
/malfunction.
• Pressure sensor 1 and/or 2 defective.
• Failure of the weighing unit / display
unit Data lines or weighing unit
defective.
• Communication error
• Malfunctioning communication
between modules.
• Communication error
• Electronics error
• Self-test calibration error
• Initial calibration not executed or
interrupted.
Scale connection not activated in the
user menu.
Error and Warning Messages
• Replace the device when error occurs
permanently.
• Replace device.
• Pump without balancing still to be
used for diagnosis.
• Replace device.
• Replace device.
•New calibration required.
Activate the scale connection in the
user menu item P02.