NOTE
After this, follow the instructions on the display.
Tool parameters that are checked by in-the-field calibration and electronically adjusted:
Target axis errors
VA collimation errors
Inclination errors of the tilt sensor (compensator)
Axis errors of the automatic prism targeting system (prism tracker)
en
NOTE
Laser pointer to cross hairs alignment error can be checked in the field. If the error is excessive, please contact Hilti
Service or, respectively, the Hilti Repair Service, as this error requires mechanical correction.
As the Hilti total station system is primarily used for measuring applications in one position, it is advisable to calibrate
the tool in the field or on the construction site at regular intervals. This is especially the case when steeper sightings
are used frequently.
7 Hilti Service
7.1 Scope
Hilti Service checks the device and, if deviations from
the specified accuracy are found, recalibrates it and
checks it again to ensure conformity with specifications.
The service certificate provides written confirmation of
conformity with specifications at the time of the test.
8 Care and maintenance
NOTE
Have damaged parts replaced by Hilti Service.
8.1 Cleaning and drying
Blow any dust off the glass.
CAUTION
Do not touch the glass surfaces with your fingers.
Use only a soft, clean cloth to clean the tool. If necessary,
the cloth may be moistened with a little pure alcohol or
water.
CAUTION
Do not use liquids other than alcohol or water. Other
liquids may damage plastic parts.
NOTE
Have damaged parts replaced by Hilti Service.
8.2 Storage
NOTE
Do not put the tool into storage when wet. Allow it to dry
before putting it away.
24
7.2 Recommended service interval
Have the device checked at suitable intervals, depending
on how heavily it is used.
Have the device checked by Hilti Service at least once a
year.
Have the device checked by Hilti Service if it has been
subjected to exceptional stress or unusually harsh con-
ditions.
Have the device checked by Hilti Service before using it
for particularly important work.
NOTE
Having the device checked by Hilti Service does not
relieve the user of his/her obligation to check the device
before and during use.
NOTE
Always clean the tool, its transport container and ac-
cessories before putting them into storage.
NOTE
Check the accuracy of the equipment before it is used
after a long period of storage or transportation.
CAUTION
Remove the batteries if the tool is to remain unused for
a long period of time. Leaking batteries may damage the
tool.
NOTE
Observe the specified temperature limits when storing
your equipment, above all in winter and summer, es-
pecially if the equipment is stored in a motor vehicle
(-30°C to +70°C (-22°F to +158°F)).