GUIDANCE AND MANUFACTURER'S DECLARATION
Electromagnetic emissions
The « KINETEC Spectra Essential » is intended for use in the electromagnetic environment specified below. The customer or
the user of the « KINETEC Spectra Essential » should ensure that it is used in such an environment.
Emissions test
Radio electric-Frequency emissions
CISPR 11
Radio electric-Frequency emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / Flicker emissions
IEC 61000-3-3
Electromagnetic immunity
The « KINETEC Spectra Essential » is intended for use in the electromagnetic environment specified below. The customer or
the user of the « KINETEC Spectra Essential » should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast transient
/ burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage interruptions
CEI 61000-4-11
Voltage dips and
voltage variations on
power supply input
lines
CEI 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE : U
is the a.c. mains voltage prior to application of the test level.
T
Spectra Essential
Compliance
Group 1
Class B
Class A
Complies
IEC 60601
Compliance
Test level
±2 kV, ±4 kV, ±6 kV
Complies
contact
±2 kV, ±4 kV, ±8 kV
Complies
air
±2 kV for power
Complies
supply lines
±1 kV between lines
Complies
±2 kV between line
and earth
< 5% U
T
Complies
(>95% dip in U
)
T
for 5 seconds
< 5% U
T
(>95% dip in U
)
T
Complies
for 0,5 cycle
40% U
T
Complies
(60% dip in U
)
T
for 5 cycles
70% U
T
Complies
(30% dip in U
)
T
for 25 cycles
Complies
3A/m
Electromagnetic environment – guidance
The « KINETEC Spectra Essential » uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment
The « KINETEC Spectra Essential » is suitable for use in all
establishments including domestic establishments and those directly
connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Electromagnetic environment – guidance
level
Floors should be wood, concrete or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
If the movement stops, push the start button, the movement
starts again.
If the yellow LED is lighting, switch off then on the apparatus and
push the start button, the movement starts again.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
At the interruption, there is a reset of the « KINETEC Spectra
Essential ». After turning on, push START to begin the session.
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the « KINETEC Spectra
Essential» requires continued operation during power supply
interruptions, it is recommended that the « KINETEC Spectra
Essential» be powered from an uninterruptible power supply or a
battery.
Power
frequency
characteristic of a typical location in a typical commercial or
hospital environment.
EN
magnetic
fields
should
be
at
levels
9/9