As with all KINETEC systems, KINETEC C.E.M. is equipped with a START/STOP/REVERSE function.
When the unit is running, the display reads RUN
Press the
Press the
RUN.
For optimum safety, always give the hand control to the patient before starting the system.
Set up the patient and proceed as below:
Beginning
Switch the unit ON
Start the session with the original parameters of the
movement (default setting).
• Treatment Mode
• Shoulder positioning
• Extension limit
• Flexion limit
• Speed
• Force
• Extension pause
• Flexion pause
• Timer
key of the hand control. The movement stops. The display reads STOP.
key of the hand control. The movement starts in the opposite direction and the display reads
Keys to press
10
1 to 5
0 to 900 seconds (15minutes)
0 to 900 seconds (15minutes)
No timer(00H00) to 24H00
Caution:
Display
KINETEC
ELBOW MODULE 1.0
Please wait
ELBOW MODULE
0
STOP
80
135
OU
WARM UP
0
STOP
80
135
ELBOW MODULE
RUN
10
135
Possible values
60° to 90°
-10 to 130°
-5° to 135°
(from 50° to 140° per minute)
1 to 6
Use of the hand control

START/STOP/REVERSE function

Quick Start

Remarks
The value change at the speed of the
movement.

Possible values for each parameter

Default setting
Normal
-
-10°
135°
2
6
0
0
0
7