ELECTRICAL SPECIFICATIONS
Power supply
Range of voltage permitted
Average battery life
DISPLAY
Display
Low battery indicator
Overload indicator
Indicator of lead reversal
Indicator of current circuit fault
Indicator of voltage circuit fault
I Resistance too high
I Interference noise
MECHANICAL SPECIFICATIONS
Protection index
Free drop
Shock resistance
Resistance to impacts
Vibration resistance
Dimensions (W x H x D)
Connection of leads
8 batteries R6 alkaline 1.5 V or 8 accumulators
8.8...12.5 V
1800 measurements of 15 seconds (at 23°C ± 3 K, with 8
alkaline batteries) permanent test of remaining battery life.
LCD, 7 segments, height 18 mm,
2000 counts (3½ digits), direct reading in Ω
"LO BAT" message
display of
1 ---
- sign preceding the measurement
by flashing red FAULT (E -H) LED
by flashing red HIGH RESISTANCE (E - S) LED
by flashing red HIGH NOISE (E - S) LED
IP 54 in accordance with IEC 529 and NF EN 60529
25 cm in accordance with IEC 68-2-32
50 g
(g
= 9.81 m/s²) in accordance with IEC 68.2.27
n
n
IK 04 in accordance with NF EN 50102
± 1 mm, from 10 to 55 Hz in accordance with IEC 68.2.6
238 x 136 x 150 mm
on 3 colour coded terminals for φ 4 mm banana plugs or
6 mm spade clips
21