Specification
Sample pressure, flow rate, and
temperature
Number of sample streams
Certifications
Specification
Light source
Measurement range
Accuracy
Precision/Repeatability
Response time
Stabilization time
Calibration time
Minimum detection limit
Reagent usage
Standard usage and cleaning
solutions

General information

In no event will the manufacturer be liable for direct, indirect, special, incidental or consequential
damages resulting from any defect or omission in this manual. The manufacturer reserves the right to
make changes in this manual and the products it describes at any time, without notice or obligation.
Revised editions are found on the manufacturer's website.

Safety information

The manufacturer is not responsible for any damages due to misapplication or misuse of this product including,
without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent
permitted under applicable law. The user is solely responsible to identify critical application risks and install
appropriate mechanisms to protect processes during a possible equipment malfunction.
Please read this entire manual before unpacking, setting up or operating this equipment. Pay
attention to all danger and caution statements. Failure to do so could result in serious injury to the
operator or damage to the equipment.
Make sure that the protection provided by this equipment is not impaired. Do not use or install this
equipment in any manner other than that specified in this manual.
4 English
Table 1 General specifications (continued)
Details
Pressure: 17.2 to 600 kPa (2.5 to 87 psi) to preset pressure regulator
Flow rate: 100 to 1000 mL/minute
Temperature: 5 to 50 °C (41 to 122 °F)
1 or 2 with programmable sequence
ETL certified to UL and CSA standards, CE marked
Table 2 Measurement specifications
Details
LED (light emitting diode) with a peak wavelength of 650 nm
0.01 to 2 ppm as N, 0.05 to 10 ppm as Cl
± 5% or 0.01 ppm as N or 0.05 ppm as Cl
(41 to 104 °F)
±10% or 0.02 as N or 0.10 ppm as Cl
(104 to 122 °F)
3% or 0.01 ppm as N or 0.05 ppm as Cl
Within 5 minutes
After initial startup or annual maintenance: 10 measurement cycles
After standby: 1 measurement cycle
After calibration: 1 measurement cycle
Automatic slope and offset calibration: 50 minutes
0.01 ppm as N or 0.05 ppm as Cl
Usage: 1 L of reagent every month with a 5-minute cycle time
Container: 1 L, HDPE with polypropylene caps
Usage: 2 L of each standard and cleaning solution monthly
Container: 2 L, PETE with polypropylene caps
N O T I C E
2
(the larger value) for 5 to 40 °C
2
(the larger value) for 40 to 50 °C
2
(the larger value)
2
2