IBP1 ALM LMT ERR
IBP2 ALM LMT ERR
Prompt message (general alerts):
Message
BP1 SYS EXCEED
IBP1 DIA EXCEED
IBP1 MEAN EXCEED
IBP2 SYS EXCEED
IBP2 DIA EXCEED
IBP2 MEAN EXCEED
IBP1 NEED ZERO-CAL
IBP2 NEED ZERO-CAL

15.6 Maintenance and Cleaning

15.6.1 Care and Cleaning
Warning
• Before cleaning the monitor or the transducer, make sure that the equipment is switched off and
disconnected from the power line.
Cleaning of IBP Transducer (Reusable)
After the IBP monitoring operation is completed, remove the tubing and the dome from the transducer
and wipe the transducer diaphragm with water. Soaking and/or wiping with soap can clean the transducer
and cable and water or cleaning agents such as those listed below:
Cetylcide
Wavicide-01
Wescodyne
Cidex
Lysol
Vesphene
Do not immerse the connector in any liquid. After cleaning, dry the transducer thoroughly before storing.
Slight discoloration or temporary increase of surface stickiness of the cable should not be considered
abnormal If adhesive tape residue must be removed from the transducer cable, double seal tape remover
is effective and will cause a minimum of damage to the cable if used sparingly. Acetone, Alcohol, Ammonia
and Chloroform, or other strong solvents are not recommended because over time the vinyl cabling will be
damaged by these agents.
Note
• The disposable transducers or domes must not be re-sterilized or re-used.
• For protecting environment, the disposable transducers or domes must be recycled or disposed of pro-
perly.
Functional safety failure
Functional safety failure
Cause
Systolic measuring value of channel 1 is
beyond measurement range.
Diastolic measuring value of channel 1 is
beyond measurement range.
Mean measuring value of channel 1 is beyond
measurement range.
Systolic measuring value of channel 2 is
beyond measurement range.
Diastolic measuring value of channel 2 is
beyond measurement range.
Mean measuring value of channel 2 is beyond
measurement range.
Zero calibrating must be done before measu-
ring in IBP channel 1.
Zero calibrating must be done before measu-
ring in IBP channel2.
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
HIGH
Stop using measuring function of IBP
module, notify biomedical engineer or
Our service staff.
Alarm Level
HIGH
HIGH
HIGH
HIGH
HIGH
HIGH
LOW
LOW