Switching to a different type of gas
The safety burner conforms with the regulations set out in DIN 30665 Part 1 and enables a switch to ano-
EN
ther type of gas by changing the set of burner nozzles.
Type of gas and power:
Type of gas
Liquid gas
G 30 / 50 mbar
Natural gas E
G 20 / 20 mbar
• The natural gas E version is distinguished with the sticker „Erdgas E 20 mbar".
• The liquid gas version is distinguished with the sticker „Flüssiggas 50 mbar".
The predetermined lab burner can be switched over to natural gas E and liquid gas by changing the set of
nozzles and its respective sticker. The relevant conversion sets are given in the table above.
Each set consists of a small nozzle, a large nozzle and a sticker naming the respective type of gas. The
conversion sets are featured in the spare parts list.
The changeover of the nozzles must only be carried out by
qualified specialists in accordance with DVGW regulations (German legislation - check local regu-
lations).

10 Technical data

Height (depending on the angle):
(with burner base):
Mode of operation:
Ambient temperature, max.:
Small nozzle Ø
0,18 mm
[7.1 mil]
0,32 mm
[12.6 mil]
Weight
Connection hose:
Supply pressure:
Consumption:
Output:
Dimensions
(W x H x D):
Rated power
Large noz-
Load
zle Ø
0,34 mm
0,75 kW
[13.4 mil]
0,55 mm
0,65 kW
[21.7 mil]
933-0100
Liquid gas
According to DIN 30664
50 mbar
48 g/h
[1.69 oz/h]
750 W
Continous operation
[6.30 x 4.80 x 2.36 inch]
- 4 -
Load large
Load small
tube
tube
0,60 kW
0,15 kW
0,50 kW
0,15 kW
Type
932-0100
Natural gas E
113 / 122 mm
[4.45 / 4.80 inch]
640 g
[22.4 oz]
20 mbar
63 l/h
[2.2 ft³/h]
650 W
40 ºC
[104 ºF]
160 x 122 x 60 mm
Product ref.
mark
933-0100
932-0100