Please contact HK Surgical for assistance if the following solutions do not resolve the problem.
PROBLEM
Pump not operating
Foot switches will not
activate pump.
Inadequate flow when
foot pedal is pressed.
Tubing continues to drip
after pump is stopped.
UMKIP Rev. N
I
N
G
G
U
I
D
E
T
I
N
G
G
U
I
D
E
POSSIBLE CAUSES
1. Power switch not on.
2. Power cord not plugged into
wall outlet.
3. Power cord not plugged into
unit.
4. Foot switch not activated.
1. Foot switch tubing not
connected.
2. Foot switch broken or leaking
3. Power switch not on.
1. Speed control set too low.
2. Wrong tubing in pump.
1. Gravitational flow
SOLUTION
1. Turn power switch to "on."
2. Plug power cord into outlet.
3. Plug power cord into connector in
rear of unit.
4. Depress foot switches.
1. Connect foot switch tube to
connector in front of power unit.
2. Check for tear or broken tubing.
Replace foot switch.
3. Turn power switch to "on."
1. Turn speed control clock-wise to
increase speed.
2. Press foot switch to turn on.
3. Only use tubing specified and
approved by HK Surgical.
4. Turn speed control to 5 and then dial
back down to desired setting.
1. Hang the IV bag level with the pump.
- 8 -