Maintain the following times when disinfecting:
Wattage
850 – 950
600 – 850
Information about the amount of water required and about the
disinfection process is included in the instructions for use of the
devices in question (disinfector and microwave).
Chemical cleaning with disinfection
Cleaning and disinfection can be carried out in a single cycle
using a chemical preparation process. For this, use the clean-
ing disinfecting agent Bomix
To ensure safety when handling chemicals, follow the instruc-
tions for use of the disinfecting agent, particularly the accompa-
nying safety instructions.
CAUTION
Risk of infection due to growth of bacteria
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Observe the mixing proportions indicated and the spe-
cified treatment time.
• Ensure that all components are completely submerged in
the solution for the whole of the treatment time. There
must not be any air pockets or bubbles.
• Prepare a 2% solution of
®
Bomix
 plus by mixing 10 ml of
the concentrate with 500 ml
tap water.
• Place all the individual parts in
the prepared solution and
leave them to soak for 5 min.
– 75 –
Disinfection time
4 min.
6 min.
®
 plus.
Cooling time
2 min.
2 min.