SYSTEM CONFIGURATIONS
The Automated Impella Controller connects to the Impella Catheter and
accessory components in the initial set-up configuration as illustrated.
Glucose
Solution
Automated Impella
®
Controller
Connector Cable
Purge Cassette
Y Connector
IMPELLA
CATHETER
®
The Impella Catheter is an intravascular microaxial blood pump. The Impella
2.5 delivers up to 2.5 liters of blood per minute from the left ventricle into the
aorta and the Impella CP delivers up to 4.1 L/min.
Impella 2.5
®
Catheter Shaft
Red Pressure
Sidearm
Clear Sidearm
Repositioning Sheath
Red Impella
®
Plug
Impella CP
®
Catheter Shaft
Red Pressure
Sidearm
Clear
Sidearm
Red Impella Plug
Reaccess Sheath
4
Impella
®
Catheter
Pigtail
Inlet Area
Radiopaque Marker
Easy
Outlet
Guide Lumen
Area
Cannula
Check Valve
Motor
Housing
Open Pressure Area
Pressure
Reservoir
Infusion Filter
Pigtail
Inlet Area
EasyGuide
Radiopaque
Lumen
Marker
Outlet
Area
Cannula
Motor
Housing
Open Pressure Area
Check Valve
Pressure
Reservoir
Infusion Filter
PURGE CASSETTE
Do not use saline in the purge system.
The purge cassette delivers rinsing fluid to the Impella Catheter. The purge
fluid (typically 5% glucose solution in water with heparin) flows from the
purge cassette through the catheter to the microaxial blood pump to prevent
blood from entering the motor. When the purge cassette is properly installed
in the Automated Impella Controller, the Abiomed
facing you. Note: The Y connector attached to the purge tubing is used for
the initial set-up configuration of the Impella 2.5
Switch to the standard configuration as soon as practical.
ACCESSORIES
Component
®
logo is upright and
®
and Impella CP
®
Systems.
Description
The white connector cable connects the
Impella Catheter to the Automated Impella
Controller. Clips on the cable are used to
secure the purge tubing to the cable.
• The socket at the grey end of the cable
connects to the Impella Catheter plug.
• The white plug at the opposite end of the
cable is inserted into the blue catheter
plug on the front of the Automated
Impella Controller.
The Impella 2.5 introducer kit is used to
gain arterial access for the Impella 2.5
Catheter. It contains:
• Introducer—with hemostatic valve for
tight fit around components and single-
step "break-away" configuration
• Dilator—easy to insert and remove with
soft design for atraumatic approach into
femoral artery
• 18 G Seldinger needle
• 12 cc syringe
• 0.035 inch stiff access guidewire