Technical data
Measurement principle: oscillometric
Measurement range:
40 to 240 mmHg
40 to 200 pulses/min
Memory store: 280 measurements
Display: LCD
Operating conditions:
ambient temperature: 10°C – 40°C
relative humidity: 10 % – 85 %
Transport and storage conditions:
Temperature -5°C – 50°C
relative humidity: max. 85 %
Power supply:
DC 3V (2 x 1.5 V microcells alkali manganese,
IEC LR 03)
Battery check:
indicator symbol in display
Weight: 82 g without batteries
Dimensions (W x D x H): 70mm x 55mm x
70mm
Classification: Type BF
Clinical Test (DIN 58130):
Accuracy complies with EN 1060 part 3
Maximum deviation of cuff pressure: ± 3 mmHg
Maximum deviation of pulse rate: ± 5%
The right to alter technical specifications is reser-
ved.
35