13. EMC / technical description
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Note: U
is the a.c. mains voltage prior to application of the test level
T
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 4)
Electromagnetic immunity
The Symphony electric breast pump is intended for use in the electromagnetic environment specified
below. The customer or the user of the Symphony electric breast pump should ensure that it is used
in such an environment.
Immunity
test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
240
3 A/m
IEC
Com-
60601
pliance
test level
level
3 Vrms
150 kHz to
80 MHz
3 Vrms
3 V/m
3 V/m
80 MHz to
2.5 GHz
Power frequency magnetic fields
should be at levels
3 A/m
characteristic of a typical location
in a typical commercial or hospi-
tal environment.
Electromagnetic
environment – guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Sym-
phony electric breast pump, including cables, than
the recommended separation distance calculated
from the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 0.35√P
d = 1.2√P
80 MHz to 800 MHz
d = 2.3√P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determi-
ned by an electromagnetic site survey,
than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
a
should be less
b