Vacuson 60LP
•
The insertion cannula must be on the left side (suction side) of the peristaltic
pump, insure that "pressure" tube runs to infiltration needle, and "suction"
tube runs to bottle with tumescence liquid bottle!
•
Use the designed tubing set (Art.-No. 6022) only!
5)
Stick insertion cannula into the Infiltration liquid bottle and hang it on the hanger rod
6)
Open Tube Clamp
7)
Open Air Valve of Clamp assembly.
8)
Install Infiltration Needle into other end of Tube.
4.6 Suction Assembly Connection
1)
With Lid firmly secured, Hang Secretion Jar in support mounting.
2)
Affix angled short piece of connecting pipe (1) onto Secretion Lid neck.
3)
Insert bacterial filter (3) on socket of main unit.
4)
Connect tube of baccterial filter (2) with bacterial filter and
angled short piece of connecting pipe.
5)
Hang Quiver on stand holder.
Insure that the VACUSON 60LP is never operated without the Bacterial Filter,
otherwise the equipment could be destroyed!
1
6)
Press handpiece (2) firmly on motor (3), until handpieces snaps in. Check fitting.
7)
Pull motor plug and motor cable (4) through bow (5) and lead bow up to the electronicmotor
8)
Insert handpiece and electronicmotor sufficiently into the handle (1).
9)
Screw tubing adapter (6) on handpiece.
10) Connect tubing adapter and handle with the short piece ot silicon tube (7).
11) Tighten bow on handle.
12) Slip suction tube over connector of the handle.
13) Connect Motor Cable to Motor socket at the control unit.
14) Affix Overflow Connector to Secretion Jar Lid ( see picture).
15) Affix Suction Tube to connector neck.
16) For ease of handling, clip Suction Tube and Motor Cable together.
17) Suction Cannula (not included) can now be installed.
6
2
3
8
2
1
4
6
3
5
7