EQUIPMENT
INSTALLATION
GENERAL
PRECAUTIONS
FOR THE
INSTALLATION
HYDRAULIC
CONNECTIONS
Connections for the
Milldrop
12
To ensure the proper working and/or prevent dangerous situations, follow these
precautions:
– Do not install
Millrack in a place that is excessively humid or poorly ventilated
– Do not install
Millrack were there is gas or inflammable and/or explosive vapors
– Arrange the power cord and the external fill/drain tubing so that they are
along their complete route.
bent or crushed
H
AVE THE ELECTRICAL CONNECTIONS AND PLUMBING PERFORMED BY
QUALIFIED TECHNICIAN
According to the equipment installed on Millrack, carry out the following hydraulic
connections.
The demineralizer Milldrop must be installed on the top space (Shelf #1).
Use the content of the supplied kit (Ref. 5 – page 4) and carry out the connections as
shown in the figure. Fasten by clamp the hose connected to the hose-union fitting
Start-Stop cable
(to sterilizer)
1.
To connect the 90-degree fittings (Ref. 3 and 7), first remove the safety ring
(Ref. 5), connect and lock them by the ring.
2.
Similarly, to plug a silicone hose into a connector first remove the safety ring,
plug-in the hose and then lock it again by the ring.
3.
Use a Teflon tape (ref. 4) on the thread's 90-degree fitting (Ref. 3), and a seal
(Ref. 1) between the adapter (Ref. 2) and Milldrop's nipple (Ref. 8).
T
O PLUG HOSE OUT A CONNECTOR REMOVE THE SAFETY RING AND PRESS ON
THE FLOWING NUT WHILE DETACHING THE HOSE
A
To plug-in the hose
NOTE
.
8
4
1
2
3
6
5
7
5
NOTE
.
B
To plug-out the hose
;
;
not