Maintenance; Storage; Troubleshooting - Metrohm Spectrosense Instrucciones Para El Uso

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

4 Maintenance

4 Maintenance

4.1 Storage

• Rinse the Spectrosense with dist. water between the titrations.
• Rinse the Spectrosense thoroughly with dist. water and dry it before
• Use only soft tissues to wipe the mirror surface and make sure no for-

4.2 Troubleshooting

Problem
There is no
output signal.
The output signal
is noisy or
unstable.
30
storing it.
eign material is on it, as this would involve the danger of the surface
getting scratched.
Remedy
• Connect the Spectrosense to the MSB connection
• Connect the Spectrosense to the measuring input
• Remove the Spectrosense from the solution and
• Check if the mirror is tightly screwed onto the
• Check if there are any foreign materials or air
• Check whether the connection to the power supply
of your Titrando or to the stirrer connection of your
Titrino.
'Ind.' of your Titrando or Titrino.
check whether there is light at the tip. If you cannot
see any light you have to replace the Spec-
trosense. The LED cannot be exchanged.
Spectrosense.
bubbles in the optical path.
exists. The connection must not be under me-
chanical stress.
Spectrosense, Instructions for Use

Publicidad

Tabla de contenido
loading

Tabla de contenido