Gima OXY-100 Guía De Uso página 38

Ocultar thumbs Ver también para OXY-100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

6.2 Alert sound mute setting
- During the measurement, if the alert sound is set to on, short time press "
" recall/mute key, then the alert sound will mute for 90 seconds, but the
over-limited value still keeps blinking. At this moment, the sound indication
icon " " becomes "
alert sound will be activated again.
- During the measurement, if the probe is off or disconnected, the message
"Check Probe" shows on the display screen. The alert sound starts (interval is
5 seconds) and lasts for about 1 minute. If the probe is still off, the Oximeter
will power off automatically.
7 Packing List
1. An Oximeter
2. A SpO2 probe
3. A holder
4. Battery ( AA ) × 3
5. Charging cable (optional)
6. User Manual
7. Quality Inspection Certificate
8. A data cable (optional)
Note: The accessories are subject to change. See the Packing
List for detailed items and quantity.
8 Repair and Maintenance
8.1 Maintenance
The expected service life(not a warranty) of this device is 5 years. In order to
ensure its long service life, please pay attention to the maintenance;
- Please change the batteries when the low-voltage indicator appears.
- Please take out the batteries if the oximeter will not be used for a long time.
- The recommended storage environment of the device:
Ambient temperature: -20ºC ~60ºC
Relative humidity 10%~95%
Atmospheric pressure: 50kPa~107.4kPa
- The oximeter is calibrated in the factory before sale, there is no need to
calibrate it during its life cycle. However, if it is necessary to verify its accuracy
routinely, the user can do the verification by means of SpO
be done by the local third party test house.
ENGLISH
". If this alert event persists over 90 seconds, then the
simulator, or it can
2
38

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido