7-4 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the handpiece and spray for approximately 2-3
seconds. Apply lubricant until it expels from the handpiece head for at least 2 seconds. (Fig. 10)
CAUTION
• When applying spray be sure to hold the handpiece firmly to prevent the handpiece from slipping out of the hand due to the spray pressure.
• Hold the spray can upright.
E-Type Spray Nozzle
Handpiece
Fig. 10