Defender 6000 Indicators
4.9.2 Input
The input I/O's function is defined below.
The input connection is disabled.
OFF
The external input initiates a Zero function
ZErO
The external input initiates a Tare function
tArE
The external input initiates a Clear Tare function
CLr.tAr
The external input initiates a Print function
PrINt
The external input initiates a Unit function
UNIt
The external input initiates an Accumulate function
ACCUmM
The external input initiates a Start function
StArt
The external input initiates a Reset function
rESEt
The external input initiates a Start function, the second external input initiates a Stop
StA-SP
function.
The external input initiates a pause function, the second external input initiates a
PA-rE
resume function.
4.9.3 Output
The output connection is disabled.
OFF
If the weight is overloaded, the instrument will perfom a relay output.
OU.LOAD
If the weight is underloaded, the instrument will perfom a relay output.
UN.LOAD
If the weight is above the over value you set, the instrument will perfom a relay output.
over
If the weight is below the under value you set, the instrument will perfom a relay
UNDEr
output.
If the weight is within the acceptable range you set, the instrument will perfom a relay
ACCEPt
output.
The weight value is below the under value or above the over value you set, the
UN-OU
instrument will perfom a relay output.
The weight value is >=SP1, the instrument will perfom a relay output.
F.SP1,
The weight value is >=SP2, the instrument will perfom a relay output.
F.SP2
The weight value is >=SP3, the instrument will perfom a relay output.
F.SP3
The weight value is >=SP4, the instrument will perfom a relay output.
F.SP4
4.10 Lock Key Configuration
L.O.C.k
This menu
is used to lock access to certain keys. When you select ON for one selection, the associated
key press will be ignored.
If you select Lock All Keys, you will lose function of all keys. To unlock, long press the Menu key for 30
seconds to enter the settings to unlock it.
Item
[L.ALL]
Lock All Keys
[L.OFF]
Lock Off Key
[L.ZErO]
Lock Zero Key
[L.PrINt]
Lock Print Key
[L.UNIt]
Lock Unit Key
[L.mModE]
Lock Mode Key
[L.mMeNU]
Lock Menu Key
[L.tarE]
Lock Tare key
[L.tArGE]
Lock Target key
[L.NUmM]
Lock Numeric Key
Reset
Note: If the Menu key has been locked, long press the Menu key for 30 seconds to enter the settings to unlock
it.
Available Settings
(bold is the default settings)
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
OFF, ON
NO, YES