3.2. Remote control commands
78
3.2.2.12. Mode.Parameter.SET1.Stop.Type
Mode.Parameter.SET1.Stop.Drift
Mode.Parameter.SET1.Stop.Time
Mode.Parameter.SET1.Stop.StopT
Parameters for SET: Type and size of the stop criterion of the titration.
.Type: Type of stop criterion after stop drift or switch-off delay time.
.Drift: Stop drift in µL/min. Applies when "drift" has been selected.
.Time: Switch-off delay time in s. Applies when "time" has been selected. "inf"
means infinite.
.StopT: Stop time in s. Applies when "time" has been selected and the value of
.Time is set to "inf.".
3.2.2.13. Mode.Parameter.TitrPara.Direction
Parameters for SET: Titration direction.
"auto" means the titration direction is determined automatically by the instru-
ment. If 2 EP's have been set in a SET titration, the titration direction is given
by the two EP's. The entry of the titration direction is then invalid.
3.2.2.14. Mode.Parameter.TitrPara.XPause
Parameter for SET: Pause time 1 in s. Is waited off before the dispensing of
the start volume.
3.2.2.15. Mode.Parameter.TitrPara.StartV.Type
Mode.Parameter.TitrPara.StartV.V
Mode.Parameter.TitrPara.StartV.Factor
Mode.Parameter.TitrPara.StartV.Rate
Parameters for SET: Start volume.
If an absolute start volume (abs.) has been selected, the volume in mL is
valid.
A relative start volume (rel.) is dispensed as a function of the sample size:
Start volume in mL = smpl size
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.16. Mode.Parameter.TitrPara.Pause
Parameter for SET: Pause time 2 in s. Is waited off after the dispensing of the
start volume.
3.2.2.17. Mode.Parameter.TitrPara.ExtrT
Parameter for SET: Extraction time in s.
3.2.2.18. Mode.Parameter.TitrPara.MeasInput
Mode.Parameter.TitrPara.Ipol
Mode.Parameter.TitrPara.Upol
Mode.Parameter.TitrPara.PolElectrTest
Parameters for SET:
factor
*
drift, time
1...20...999
0...10...999, inf
0...99 999, OFF
+, -, auto
0...999 999
abs., rel., OFF
0...999.99
0...±999 999
0.01...150, max.
0...999 999
0...999 999
1, 2, diff.
-127...1...+127
-1270...400...+1270
ON, OFF
719 S Titrino