SEITE 1 - 148 x 210 mm - 3100021788/01 - SCHWARZ - 20-052 (kn)
ENGLISH
en
Elipar
™
DeepCure-L
LED Curing Light
Table of Contents
Handpiece
from/into the Handpiece
3M Adhesives
Safety Instructions
PLEASE NOTE!
Prior to installation and start-up of the device, please
read these instructions carefully.
As with all technical devices, the proper function and
safe operation of this device depend on the user's
compliance with the standard safety procedures as well
as the specific safety recommendations presented in
Page
these Operating Instructions.
1
1. Use of the device is restricted to trained personnel in
accordance with the instructions below. The manu-
3
facturer assumes no liability for any damage arising
3
from any other or improper use of this device.
3
2. The charger must be accessible at all times. Do not
3
use the charger for any use other than the charging
4
of the Elipar
™
4
handpiece from the mains by unplugging the charger
from the electrical outlet.
4
Treating patients using the handpiece while it is still
4
connected to the charger is prohibited for safety
4
reasons. Light-curing is possible only if the charger
4
has been disconnected.
4
3. Use only the charger "CPS008050100" or "GTM
5
96060-0606-1.0" (AC adapter plug) which is
5
provided with the device. The use of any other
6
charger can result in damage to the battery.
6
4. CAUTION! Avoid using this device in close proximity to
6
other equipment or with other equipment in stacked
6
form, as this may result in improper function. If use in
the manner described before is necessary, this device
6
and the other equipment should be observed carefully
7
in order to ensure that they work properly.
7
5. CAUTION! Do not stare at source. May be harmful
to the eyes. Restrict exposure to the area of the
7
oral cavity in which clinical treatment is intended.
8
Protect patient and user from reflection and intensive
8
scattered light by taking the appropriate precautions,
e.g., glare shields, goggles, or coverings.
8
6. CAUTION! As is the case for all high-intensity light-
9
curing devices, the high light intensity is accompanied
9
by heat generation on the exposed surface. This heat
9
can result in irreversible damage if there is longer
12
exposure in the proximity of the pulp or soft tissue.
The exposure times given in the manufacturer's
12
instructions must be observed exactly to avoid any
such damage.
12
Uninterrupted exposure times of the same tooth
surface in excess of 20 seconds and direct contact
12
with oral mucosa or skin must be strictly avoided.
12
Scientists working in this field are in agreement
12
that the irritation caused by heat generated during
13
light curing can be minimized by taking two simple
precautions:
DeepCure-L handpiece. Disconnect the
1