When using NSK Care3 Plus automatic handpiece cleaning and
lubrication system, refer to the Care3 System instructions.
6-3 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after
each patient and sterilize as noted below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.
Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or
3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required
for use.
CAUTION
Do not autoclave the product with other instruments even when it
is in a pouch. This is to prevent possible discoloration and damage
to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature,
humidity, ventilation, and sunlight. The air should be free from
dust, salt and sulphur.
Immediately after use, the product should be cleaned, lubricated
and sterilized. If blood remains on the external or internal surfaces
it can become clotted and cause rust.
Do not heat or cool the product too quickly. Rapid change in
temperature could cause damage to the product.
If the sterilizer chamber temperature may exceed 135°C during
the dry cycle then delete the dry cycle.
Autoclave sterilization is recommended for the product. The
validity of other sterilization methods is not confirmed.
Do not touch the product immediately after autoclaving as it will
be very hot and must remain in a sterile condition.