Aligning Stylus to Spindle Centre

1.
Loosen clamping screws "C" (Fig. 10) (2x) then slightly retighten.
2.
Adjust the probe with the adjustment screws "A" (Fig. 10) (4x) to within less than 20 µm.
3.
Tighten the clamping screws "C" (2x) a little more.
4.
Adjust the probe with the adjustment screws "A" (4x) to within less than 5 µm.
5.
Tighten the clamping screws "C" (2x) to the torque specified in Fig. 10.
6.
Tighten the adjustment screws "A" (4x) to the torque specified in Fig. 10 against each other.
7.
Check the alignment.
8.
Calibrate the probe.
A = Adjustment screw
DIN913 M4x8 AF2 mm (4x)
M=1,5 Nm (1.1 lbf.ft)
Fig. 10
Aligning stylus to spindle centre
(1352)
C = Clamping screw
AF2,5 mm (2x) (1351)
M=2 Nm (1.47 lbf.ft)
Hexagon key
AF2,5 mm (1346)
AF2 mm (1097)
Dial gauge
(3079)