Technical specifications
The mask contains passive venting to protect against rebreathing. As
Pressure–flow
curve
a result of manufacturing variations, the vent flow rate may vary.
60
50
40
30
20
10
0
4
6
8
Physical dead space is the empty volume of the mask to the end of
Dead space
information
the swivel. Using the large cushions it is 145 mL.
Therapy
4 to 20 cm H
pressure
Resistance
Drop in pressure measured (nominal)
at 50 L/min: 0.3 cm H
at 100 L/min: 0.9 cm H
Environmental
Operating temperature: +41°F to +104°F (+5°C to +40°C)
conditions
Operating humidity: 15% to 95% relative humidity non-condensing
Storage and transport temperature: -4°F to +140°F( -20°C to +60°C)
Storage and transport humidity: up to 95% relative humidity
non-condensing
Mask fully assembled - no headgear
Gross
dimensions
Large:
6.10" (H) x 3.58" (W) x 4.65" (D)
(155 mm (H) x 91 mm (W) x 118 mm (D))
Select
Mask setting
options
the mask option.
Notes:
•
The mask system does not contain latex, PVC or DEHP materials.
•
The manufacturer reserves the right to change these specifications
without notice.
10
12
14
Mask pressure (cm H
0)
2
O
2
O
2
O
2
'ACTIVA'
(if available), otherwise select
5
Pressure
(cm H
4
8
12
16
16
18
20
20
'STANDARD'
Flow
O)
(L/min)
2
19
28
34
40
45
as