Problem / Malfunc-
Possible cause
tion
No display
• No voltage
• Loose connect-
ing cable
• Display defect-
ive
D1.02
(The software ver-
sion for the display
(perman-
will appear).
ent)
• Not communic-
ating with the
controller
Error
• An error has oc-
curred
Error 01
• Controller mal-
function
Error 03
• Microcontroller
Fault / MCP2
Error 04
• EEPROM error
Error 22
• Moisture de-
tector 1 cable
break
Error 32
• Moisture de-
tector 2 cable
break
Error 40
• General error
temperature
sensor 1
Error 41
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 1
Error 42
• Excess temper-
ature / short-
circuit temper-
ature sensor 1
Error 43
• Measurement
fluctuation
temperature
sensor 1
Error 50
• General error
temperature
sensor 2
Error 51
• Low temperat-
ure / short-cir-
cuit temperat-
ure sensor 2
18
Bühler Technologies GmbH
Action
• Check the sup-
ply cable
• Check fuse
• Check connec-
tions
• Check connec-
tions
• Read the error
number as de-
scribed above
• Clear error
(temporary
fault)
• Disconnect
from power for
approx. 5 s
• Contact service
• Contact service
• Contact service
• Check moisture
detector line
• Check moisture
detector
• Check moisture
detector line
• Check moisture
detector
• Sensor possibly
defective
• Check temper-
ature sensor
connection
(Flashing)
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
• Sensor possibly
defective
• Check temper-
ature sensor
connection
Error 52
• Excess temper-
ature / short-
circuit temper-
ature sensor 2
Error 53
• Measurement
fluctuation
temperature
sensor 2
Status text
Possible cause
H2o.1
• Moisture alarm
moisture de-
tector 1
init
• Heated line ini-
tialisation/
heat-up phase
PuMP
• Pumps deactiv-
ated
dt
Active Delta T con-
trol only: The block
temperature is out-
side the defined
temperature range.
• Cooler is still in
the "break-in
phase"
• Fluctuating
ambient tem-
perature
• Insufficient
cooling capa-
city
Manu
• Gas pump was
manually activ-
ated for 30
seconds
Mode
• Heated line
Addon
with Pt100 de-
tected, regu-
lator inactive
• Excess/low
temperature
• Check temper-
ature sensor
connection
• Check temper-
ature sensor
connection
Action
• Dry
• Check condens-
ate trap
• Wait
• Reactive pumps
via menu
• Wait to see if
the target tem-
perature will be
reached
• Check the ambi-
ent temperat-
ure / power
supplied
• Process-related:
Adjust alarm
limits
• Not required
• Activate regu-
lator in the
menu
• see chapter

"Troubleshoot-

ing"