MANUAL DE INSTRUCCIONES CODIGO 80191 REV D
32050
28129
Adjusting screew
J
.P. SELECTA s.a.u.
ENE/2020
10 Technical service
For repair and substitution of deteriorated parts, please contact with J.P. SELECTA,
s.a.u. or your distributor. They will inform you about the technical services author-
ized in your area.
Find a spare parts list for easy buying and identifying.
10.1 Spare parts
16213 Electrovalve N. A.
16067 Water electrovalve.
16025 Distilled outlet clamp valve.
24278 Water bomb support.
28129 Lockgum.
28061 Drop tray.
32050 Steam generator 2400W.
47043 Cooler glass DE-1626
47044 Rectifier glass DE-1626
47045 Cooled union tube glass DE-1626
47259 Anti feedback valve.
29520 Relay electronic PCB.
29511 Control panel electronic PCB
Hoses:
46030 TEFLON TUBE 8X10MM Tube inmersed on the sample.
46041 Silicone tube 8x14
46042 Silicone tube 6x10
46143 3/4 WATER INLET HOSE Connection to the water network
28264 LFL Tygon tube Distilled outlet.
10.1.1 Adjusting the level sensor
It could be possible that along the time the collector level sensor changes it sensibility
and never detects the distilled level or it actuates just by sensing the glass flask.
Turn the adjusting screew (see picture) to adjust it.
Place a flask with water on the collector and turn the screew until the yellow LED ligths.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso)
Water and Steam.
Water and Steam.
WARNING
Before any work inside the equipment
disconnect it from the mains.
Tel 34 937 700 877 Fax 34 937 702 362
Pag.: 32