Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 176
ALARM STATUS
Critical high priority
High priority
EN
Medium priority
Low priority
IMPORTANT: If two alarm conditions exist at the same time, the highest priority alarm is
indicated. If two or more alarm conditions of equal priority exist at the same time, the
most recent one will be displayed.
IMPORTANT: The most recent alarms indicated by the device are logged for reference
by service personnel. This log is maintained even if the device is powered down or if
power is lost for any other reason.
IMPORTANT: If the mute button is pressed prior to an alarm condition (for example, to
mute the device in a movie theatre), critical high priority and high priority alarms will
override the mute function; medium and low priority alarms will be muted for eight hours
and twenty four hours respectively from the time the button was pressed. Press the mute
button off to display the last highest priority alarm. Press the mute button on again to
reset the eight-hour timer.

9 . 1 . ALARMS

When the concentrator sounds an alarm, a corresponding message will be displayed on
the screen. Take appropriate action as directed in the charts below.
AUDIBLE TONE
Ten beeps per burst,
burst repeats every 3
seconds.
Ten beeps per burst,
burst repeats every 3
seconds.
Three beeps per
burst, burst repeats
every 8 seconds
Three beeps per
burst, burst repeats
every 10 minutes
24/340
VISUAL INDICATOR
Solid red LED and
device is automati-
cally disabled
Flashing red LED
Flashing yellow LED
Solid yellow LED
MUTE TIME
20 minutes
20 minutes
8 hours
24 hours

Publicidad

Tabla de contenido

Solución de problemas

loading

Productos relacionados para GCE Zen-O lite RS-00600

Este manual también es adecuado para:

Zen-o lite rs 00600c

Tabla de contenido