13. Re-insert the battery into the device.
14. Connect the device to an AC or DC power source.
DO NOT POWER ON!
15. Press the alarm mute
shows "Leak Test".
a. The device will indicate progress by decimals right of the "Leak test" text.
b. When completed, the device will indicate "Test Passed" or "Test Failed".
i
Test Passed - Press battery button
ii
Test Failed - Press battery button
16. If the leak test has passed, then clear the device sieve hour counter.
a. Turn on your Zen-O lite portable oxygen concentrator.
b. Simultaneously press the
c. Press the
displayed, where XXXX = number of hours for the old sieve was utilised.
d. Press the alarm mute button
e. Press the battery button
f.
The device is now ready to be used.
Important: Contact your home oxygen provider or distributor if the device does not pass
after multiple attempts.
12 .3 . DISPOSAL
• Contact your distributor regarding disposal of the device.
• Dispose of battery according to local regulations or contact your distributor.
13 . WARRANTY
The device warranty is limited to three (3) years from date of sale to GCE´s customer, this
excludes the sieve modules which have only one year warranty. All accessories including
batteries are limited to one (1) year warranty.
The standard warranty is only valid for products handled as stated in the user manual and
in accordance with general industry good practice and standards.
14 . TRADEMARKS AND DISCLAIMER
14 . 1 . TRADEMARK
All trademarks are the property of their respective owners.
and battery
and
or
button to navigate he diagnostic menu until "XXXH Sieve" is
until the sieve hour counter shows "0H Sieve".
to exit the diagnostic mode.
31/340
buttons simultaneously until the displays
to exit, continue to step 16.
to exit, then repeat steps 4-15.
buttons until "Diagnostics" is displayed.
EN