Guidance and Manufacturer's Declaration: Electromagnetic Emissions
The HD Camera Control System is intended for use in the electromagnetic environment specified
below . The customer or the user of the HD Camera Control System should ensure that it is used in
such an environment
Immunity Test
IEC 60601 Test
Level
Electrostatic
±8 kV contact
Discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/
output lines
Surge
±1 kV line(s) to
IEC 61000-4-5
line(s)
±8 kV line(s) to
earth
Voltage dips,
<5 % U (95% dip
short interruptions
in U ) for 0 .5 cycle
and voltage
40 % U (60% dip
variations on
in U ) for 5 cycles
power supply
70 % U (30% dip
input lines
in U ) for 25 cycle
IEC 61000-4-11
<5 % U (95% dip
in U ) for 5 s
NOTE: U is the AC mains voltage prior to application of the test level .
Compliance Level Electromagnetic Environment
Guidance
±2,4,8kV contact
Floors should be wood, concrete,
±2,4,15kV air
or ceramic tile . If floors are covered
with synthetic material, the relative
humidity should be at least 30% .
±2kV for power
Mains power quality should be that
supply lines
of a typical commercial or hospital
±1kV for input/
environment .
output lines
±1 kV line(s) to
Mains power quality should be that
line(s)
of a typical commercial or hospital
±8 kV line(s) to
environment .
earth
<5 % U (95% dip
Mains power quality should be that
in U ) for 0 .5 cycle
of a typical commercial or hospital
40 % U (60% dip
environment . If the user of the HD
in U ) for 5 cycles
Camera
70 % U (30% dip
in U ) for 25 cycle
Control Unit requires continued
<5 % U (95% dip
operation during power mains
in U ) for 5 s
interruptions, it is recommended
that the HD Camera Control System
be powered from an uninterruptible
power supply or a battery .