NOUVAG AG
7.5 Seals and Tubes
To insure proper function of the Suction Pump, all tubes and seals must be periodically checked, and
replaced after at most 250 sterilizations or five years operation. Only when in perfect order can
sufficient vacuum be built up.
Defective seals and tubes must be replaced!
2
1. Seal for connecting neck on Control Unit ...................................................... 2 pieces ........ (Art.4063)
2. Connecting Tube between Bacteria Filter and Jar lid ...................................... 1 piece ........ (Art.4075)
3. Seal for unscrew able connecting neck on Secretion Jar Lid ........................ 4 pieces ........ (Art.4064)
4. Seal between Jar Lid and Jar .......................................................................... 1 piece ...... (Art.28957)
5. Seal between wide unscrew able connector neck and Secretion Jar Lid ..... 4 pieces ........ (Art.4063)
6. Seal in Overflow Protection .............................................................................. 1 piece ...... (Art.28958)
7. Suction Tube ............................................................ 1piece (see 9. Replacement parts and numbers)
7.6 Control Unit Fuse Replacement
A blown Fuse can be easily replaced by the operator. It is located on the back of the Control Unit:
1)
Disconnect mains.
2)
With screwdriver open fuse cover by turning counter clockwise.
3)
Replace defective fuse. (see 2.2 Technical Data
4)
With Screwdriver close fuse case by turning clockwise.
5)
Reconnect mains.
7.7 Disposal
Disposal of device, components and accessories must strictly conform to local laws and regulations as
set out by the relevant authorities.
With regard to the preservation of the environment old equipment may be returned to the distributor or
manufacturer.
No. 31791 04/08