APPENDIX (continued)
Guidance and manufacturer's declaration - electromagnetic immunity
The Integra Padgett Electric Dermatome Models B and PI are intended for use in the
electromagnetic environment specified below. The customer or the user of the Integra
Padgett Dermatome Models B and PI should assure that they are being used in such an
environment
Immunity Test
Electrostatic
discharge, (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
NOTE UT is the a. c. mains voltage prior to application of the test level.
25
®
IEC 60601 test level
±6 kV contact
±8 kV air
± 2 kV for power
supply lines
± 1 kV for input/
output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
3 A/m
<5 % UT
for 0,5 cycle
<40 % UT
for 5 cycles
<70 % UT
for 25 cycles
<5 % UT
for 5 sec
Compliance
Electromagnetic Environment -
level
Guidance
±6 kV
Floors should be wood, concrete
±8 kV
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
±2 kV
Mains power quality should be
that of a typical commercial or
±1 kV
hospital environment.
±1 kV
Mains power quality should be
that of a typical commercial or
±2 kV
hospital environment
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
><5 % UT
Mains power quality should
for 0,5 cycle
be that of a typical commercial
or hospital environment. If the
<40 % UT
user of the Integra Padgett
for 5 cycles
Electric Dermatome Models
B and PI requires continued
<70 % UT
operation during power mains
for 25 cycles
interruptions, it is recommended
that the Integra Padgett Electric
<5 % UT
Dermatome Models B and PI be
for 5 sec
powered from an uniterruptible
power supply or a battery.