11 . MAINTENANCE AND CLEANING

11 . 1 . ROUTINE MAINTENANCE

WARNING: Do not use lubricants on this device or any of its accessories.
CAUTION: Replace the cannula on a regular basis. Check with your distributor or clini-
cian to determine how often the cannula should be replaced.
Device will indicate with an alarm when service is required. (Also, see Chapter 10 - Trou-
bleshooting.)
EN
No special maintenance needs to be carried out by the user. Contact your provider or
manufacturer for help, in setting up, maintenance, or to report unexpected errors.

11 .2 . CLEANING AND DISINFECTION

WARNING: Do not submerge this device in liquid . Do not expose to water or precipita-
tion . Do not expose to dusty conditions .
CAUTION: Do not use cleaning agents other than those specified in this manual. Allow
the cleaning solution to dry from the cleaned surface before use.
CAUTION: Always disconnect power and turn off this device before cleaning.
Clean the exterior with a soft cloth slightly dampened with soapy water or with anti-bac-
terial wipes (Isopropyl alcohol 70% solution).
For disinfecting, use a MadaCide-FDW-Plus wipe or equivalent and follow the manufac-
turer's instructions. (Manufacturer - Mada Medical Products Inc., www.madamedical.com)
Important: The device should receive an external cleaning weekly, accessories should
be cleaned as needed. The device exterior and accessories should be cleaned and dis-
infected and the patient filter replaced prior to delivering to a new patient. The device
may become hot after operation, take additional care when replacing filters.
Nasal cannula: Refer to the original manufacturer's instructions for cleaning the nasal
cannula.

11 .3 . SERVICE LIFE

The expected service life of the device is 5 years, except for the sieve modules. The
service life of the sieve modules will depend on the operating conditions. The sieve mod-
ules are an internal component of the device and should only be replaced by a trained
person. If the intake and exhaust vents are not blocked and the check vents alarm is
activated contact your distributor.