OPERATION/ALARM LOG

Displaying alarm log

The incubator can save alarm log data (maximum 256 logs). Alarm log data can be displayed graphically
on the LCD touch panel.
Note: When number of saved alarm logs exceeds 256, the oldest log is deleted and overwritten.
1. Go to the Log screen (Menu
Log). Press
"Alarm" to display the Alarm screen.
2. Alarm logs from the last 7 days (including current
day) are displayed.
Note: When the number of alarm logs exceed 6,
pressing the top ( ) or the bottom ( ) log, scrolls
the log table and allows hidden alarm logs to be
seen.
3. Change the number in the "Last X Days" input
box to display alarm logs for specified days
(including current day).
Settable range: 1 day
45 days.
Note: Accurate to about 1 minute per month.
.