5. The system check starts automatically. If the
system is normal, the display goes from Step1 to
Step2.
6. Press "OK" to go to Step3.
7. At Step3, CO
gas density calibration starts. The
2
display proceeds automatically to Step5.
8. After CO
density calibration finishes, the
2
incubator returns to the normal operation. Press
"OK".
63