Remote alarm terminal

The alarm of this product can be transferred to a remote location by connecting the external alarm unit to
the remote alarm terminal. For the behaviour of remote alarm output, refer to pages 70–73.
The terminal of the remote alarm is installed at the
rear upper right of the unit (see the figure opposite).
The alarm outputs from this terminal. Contact
capacity is DC 30 V, 2 A.
Table 1 shows the behaviour of the remote alarm
when the Buzzer key is pressed.
Notes:
The door alarm does not work remotely. Refer to
pages 70 and 72.
Wiring of the remote alarm should be performed
by qualified service personnel.
Table 1. Behaviour of remote alarm when pressing Buzzer key
Remote Alarm setting
(Refer to pages 41–42)
ON: Remote alarm setting not
connected with Buzzer key
(Remote alarm is cancelled)
OFF: Remote alarm setting
connected with Buzzer key
(Remote alarm is not
cancelled)
* In the case of Err01 (CO
and Err12 (CO
sensor error), the condition returns to normal.
2
Use a twisted shield wire for the connection.
Type: UL 2343, UL 2448, UL 2464, UL 2552, UL2623.
Length: 30 m max.
Connecting
Normal
terminal
condition
COM.-N.C.
COM.-N.O.
COM.-N.C.
COM.-N.O.
gas cylinder empty) , Err02 (N
2
Abnormal condition
(Including in the cases of power outage and
of where the power supply plug is pulled out)
When pressing the buzzer key
Close
Open
Open (Maintain in abnormality)*
Open
Close
Close (Maintain in abnormality)*
Close
Open
Open
Close
gas cylinder or O
gas cylinder empty), Err11
2
2
Remote alarm
terminal
Close (Return to normal)
Open (Return to normal)