Phcbi MCO-170M Instrucciones De Operación página 71

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Alarm or safety
function
The ambient temperature sensor is
Ambient
disconnected.
temperature
The ambient temperature sensor is
sensor error
short-circuited.
CO
sensor
The Vref or Vgas output voltage for the
2
error
CO
sensor is abnormal.
2
The measured voltage
O
sensor error
2
for the O
sensor is abnormal.
2
Main heater
Main heater burnout occurs or the main
error
heater SSR is short-circuited.
Bottom heater
Bottom heater burnout occurs or the
error
bottom heater SSR is short-circuited.
Door heater
Door heater burnout occurs or the door
error
heater SSR is short-circuited.
a) High-limit temperature alarm is activated.
Sensor box
b) Sensor box heater burnout occurs or
1
heater error*
the sensor box heater SSR is
short-circuited.
a) High-limit temperature alarm is activated.
Heater SSR
b) Main, bottom, door, or sensor box
1
burnout*
heater SSR burnout occurs.
(When optional MCO-170UVS is
UV lamp failure
installed)
The UV lamp burns out.
(When optional MCO-170UVS is
New UV lamp
installed)
replacement
The accumulated ON time reaches
approx. 5,000 hours.
Communication between LCD touch
Communication
panel and control substrate is
error
terminated or unstable.
After power switch is turned ON, under
Warming-up of
warming-up before temperature is
gas control
stable and gas control is enabled.
*1: Shortly after the high-limit temperature alarm is activated, Err16 (Sensor box heater error) and Err17
(Heater SSR burnout) are activated.
Conditions
"Err09: AT Sensor Open" is displayed in
the message display field.
"Err10: AT Sensor Short" is displayed in
the message display field.
"Err11: CO2 Sensor Vref Abnormal" or
"Err12: CO2 Sensor Vgas Abnormal" is
displayed in the message display field.
"Err19: O2 Sensor Abnormal" is
displayed in the message display field.
"Err13: Main Heater Abnormal" is
displayed in the message display field.
"Err14: Humidity Heater Abnormal"
is displayed in the message display field.
"Err15: Door Heater Abnormal" is
displayed in the message display field.
"Err16: Sensor Box Heater Abnormal" is
displayed in the message display field.
"Err17: Heater SSR Open" is displayed in
the message display field.
"Err18: UV Lamp Abnormal" is displayed
in the message display field.
"Warning: UV Bulb Life" is displayed in
the message display field.
"Err56: Communication Failure" is
displayed in the message display field.
"Status: Gas sensor initializing" is
displayed in the message display field.
Display
Buzzer
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
Intermittent
tone
-----
-----
-----
Remote
Safety
alarm
operation
-----
ON
-----
ON
CO
valve is
2
ON
closed.
N
/O
valve
2
2
ON
is closed.
-----
ON
-----
ON
-----
ON
-----
ON
-----
ON
-----
ON
-----
-----
-----
-----
-----
-----
71