BASIC PARAMETERS
Setting CO
The CO
gas control mode during the normal operation of the incubator can be set as follows. The incubator
2
automatically starts operation using these settings after power-on.
Setting values: ON or OFF
Factory setting: OFF
1. On the home screen, press the menu key (MENU/HOME).
▶ The left side of the display will change to the Menu screen.
2. When the cursor on the menu screen is on Set Value [Fig. 1], press
the enter key (ENTER).
▶ The display changes to the Set Value screen.
3. On the Set Value screen, move the cursor to CO
using the up/down keys (△▽), and press the enter key (ENTER).
▶ The left side of the display changes to the CO
the current set value (ON) is displayed [Fig. 3].
4. Use the up/down keys (△▽) to change the control setting value.
Each time the up/down key is pressed the value changes between ON
and OFF.
5. Press the enter key (ENTER).
▶ If it is set to ON: the value is confirmed and the display changes to the
CO
Setting screen.
2
If it is set to OFF: the value is confirmed and the display changes to
the Set Value screen.
6. Press the menu key (MENU/HOME) to display the home screen.
The display returns to the home screen automatically when 90
seconds has passed without any key operation.
gas
2
ON/OFF [Fig. 2]
2
ON/OFF screen, and
2
Menu
Set Value
Data Log
Lock
Alarms & Controls
<Left side of the display>
Set Value
Temperature
: 37.0°C
CO
ON/OFF
: ON
2
CO
Setting
: 5.0%
2
Overtemp
: 53.0°C
<Left side of the display>
CO
ON/OFF
2
ON
ON / OFF
<Right side of the display>
[Fig. 1]
[Fig. 2]
[Fig. 3]
33