Product information
Maintenance
After 2,000 hours of operation, KINETEC Breva requires a few lubrication and maintenance operations (lubrication of the joints, pointer stops and
ball screws).
The need for maintenance is indicated by display of the message SERV. MOTOR when the system is switched on.
Despite that warning, you can continue to use your KINETEC Breva by pressing START, but you should contact your nearest KINETEC technician
to have the maintenance operations conducted as soon as possible.
• A motor running time counter is available. Simultaneously press keys
RESET TIME 924H
Reset: limit low

Cleaning

Before conducting any cleaning operation, SWITCH the unit OFF and disconnect the power supply.
Use a DISINFECTANT (PROPANOL/ISOPROPANOL or ALDEHYDE-based solution). Spray the disinfectant on the SURFACES (plastic shells and
metal components).
In order to ensure optimal hygiene, you are advised to clean the covers for each new patient. All the consumables enable hazard-free disposal.

Technical specifications

Product:
Weight: 12,5kg (27.5 lb.)
Unit dimensions:
Length 56cm (22 in.),
width 37cm (15.5 in.),
height 45cm (17.7 in.)
Angular limits: see page 6
Speed: from 50° to 150°/min
Foot sizing:
from 19cm (7.5in.) to 29cm (11.4in.)

Troubleshooting

A spare parts list and technical catalog are available to you on request from your KINETEC distributor.
If, after connecting the power supply cable to the power supply and switching on KINETEC Breva:
• The display does not indicate any information:
- Check that the electrical socket is live using another device.
- Check if the remote control is well connected.
- Replace the fuse(s) (36) of the connector with fuses of the same type and caliber:
2 fuses T 750 mA 250V (6.3 x 32) (KINETEC order: 4610007434).
- If the display still does not indicate any information, contact your nearest KINETEC
technician.
• Your KINETEC Breva does not work and the display indicates 50 STOP 25 115,
Press START again.
Your KINETEC Breva still does not function: Contact your nearest KINETEC technician.
• Your KINETEC Breva does not function and the display indicates:
SERVICE D1: angle measurement function failure,
or SERVICE D2: no movement,
or SERVICE D3: abnormal consumption,
or SERVICE D4: motor connection failure,
Contact your nearest KINETEC technician if the same message is displayed after having witched the device off, then on,
and started it by pressing START.
GB
16
this counter can be resetting by pressing the key
Electrical:
Power supply: 100-240 V~
Frequency: 50-60 Hz
Power consumption: 50 VA
Device of type B class I
IP 20.
Fuse T 750mA 250V 6.3x32mm
(KINETEC order: 4610007434)
and
, the displays shows
.
Environment
- Storage/transport conditions:
Temperature: -40 to 160°F / -40 to 70°C
Relative humidity: up to 90%
- Operating conditions:
Room temperature: 50 to 105°F / 10 to 40°C
Relative humidity: up to 80%