5. Using the ventilator

5.1 Switching on/self-test

60
50
40
30
CP R
20
10
0
Stenosis
Disconnection
< 2,7 bar O
2
60
50
40
CP R
30
20
10
0
1. Please open the valve of the oxygen cylinder
slowly
. The pressure gauge will now show the
pressure in the cylinder.
2. Where appropriate, calculate the remaining op-
erating time (see "5.13 Calculation of oxygen
content/remaining operating time" on page 39).
You should change the bottle in good time, e.g.
at below 50 bar, to ensure an adequate operat-
ing time.
3. Select the desired ventilation settings (see "5.2
Selecting the ventilation settings" on page 24).
4. Switch on the MEDUMAT Easy CPR with ON/
OFF switch. The ventilator will then run a self-
test lasting approx. 2 seconds.
If audio response is enabled, before the self-test
you will hear "Open oxygen cylinder".
During this test, the four LEDs in alarm panel flash
on and off and a short acoustic alarm sounds.
If an error is found, all the LEDs in alarm panel
continue flashing and an alarm sounds. If this
happens, MEDUMAT Easy CPR must not be used
for ventilation.
Using the ventilator
EN
23