Using the ventilator
The length of the expiration phase matches the
length of the inspiration phase (respiration-time
ratio is 1:1).
4. Alternate between undertaking chest compres-
sions 30 times and ventilation 2 times.
5. At the end of the heart and lung resuscitation,
switch CPR mode off by pressing the CPR key .
Notes:
Voice messages and metronome are constant.
We advise that you conform to the metronome
and audio responses.
If CPR mode is switched on during an active
resuscitation and the voice messages and the
metronome are not synchronized to the resusci-
tation, current guidelines recommend undertak-
ing chest compressions before ventilation.
Audio responses for the "stenosis" and "discon-
nection" alarms are deactivated during CPR
mode.
If the metronome is switched on, pause the acous-
tic alarms during the phases with chest compres-
sions and audio responses.
If the key on the MEDUtrigger is pressed and no
respiratory stroke is triggered (e.g. if the previ-
ous respiratory stroke is not yet over or if CPR
mode is not activated) then a notification alarm
sounds.
Audio responses and metronome can be
switched off (see "5.11 Audio response for user
guidance" on page 35 and "5.12 Switching the
metronome on and off" on page 39). This makes
particular sense if the MEDUMAT Easy CPR is be-
ing used with other devices that have audio re-
sponses and/or a metronome (e.g. AED).
CPR mode cannot be activated when in De-
mandflow mode. CPR mode switches itself off as
soon as Demandflow mode is activated.