Hygiene Preparation; Medumat Easy Cpr; Patient Valve - Weinmann MEDUMAT Easy CPR Descripción Del Aparato E Instrucciones

Ocultar thumbs Ver también para MEDUMAT Easy CPR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Hygiene preparation

6.1 MEDUMAT Easy CPR

6.2 Patient valve

EN
42
Hygiene preparation
After every use
the MEDUMAT Easy CPR and any ac-
cessories used must undergo hygienic preparation.
Be sure to carry out a functional check after ev-
ery hygienic preparation (see "7. Functional
check" on page 47).
This product may contain disposable items.
posable items are intended to be used only once.
So use these items only once and do
cess them. Reprocessing disposable items may
impair the functionality and safety of the prod-
uct and lead to unforeseeable reactions as a re-
sult of ageing, embrittlement, wear, thermal
load, the effects of chemical processes, etc.
You can keep MEDUMAT Easy CPR clean by simply
wiping with disinfectant as described in section 6.7.
Never immerse MEDUMAT Easy CPR in disinfectants
or other liquids. Otherwise damage to the device
and thus a hazard to users and patients may result.
Always grasp the hoses by their ends. Otherwise you
might damage or tear them.
1. Disconnect the patient valve from the hoses.
Dis-
not
repro-

Publicidad

Tabla de contenido

Solución de problemas

loading

Tabla de contenido