6.7 Method
Parts
MEDUMAT Easy CPR
Patient valve
Silicone ventilation
mask
Ventilation hose
Hose casing, reusable
Oxygen fittings
MEDUtrigger
Never immerse the fittings in disinfectants or other
liquids. You may only disinfect them by wiping. On
no account may liquid get into the pressure reducer,
as this could cause explosions.
Hygienic preparation of the MEDUMAT Easy CPR
and the accessories used should be performed as de-
scribed in the following table.
Follow the instructions for use for the disinfectant
used. We recommend gigasept
fecting by immersion and terralin
fecting by wiping. We recommend the use of
suitable gloves (e.g. household or disposable gloves)
for disinfection.
Cleaning
With a dry or
Disinfect by wiping
damp cloth
Immerse in dilute solution so
that all surfaces, inside and
out, are thoroughly wetted
in hot water
without bubbles. Wait until
with a mild
the full exposure time has
detergent
elapsed. After disinfection,
rinse all parts thoroughly
inside and out with distilled
water and leave to dry.
Wipe with a
Rinse cycle 30°C,
dry or damp
no spin
cloth
With a dry or
Disinfect by wiping
damp cloth
With a dry or
Disinfect by wiping
damp cloth
Disinfection
(1)
(2)
Hygiene preparation
®
FF (new) for disin-
®
protect for disin-
Thermal
Sterilization
disinfector
Not permitted
Not permitted
Steam steril-
ization at
Cycle up to
134 °C in de-
95 °C (thermal
vices to
EN 285, resi-
disinfection in
dence time at
an automatic
least 5 - 18
cleaning
minutes.
machine)
Possible during
Not permitted
cycle
Not permitted
Not permitted
Not permitted
Not permitted
EN
45