7.3 Checking for leaks in the system

Always keep a stock of
washers for the connec-
tions available.
Important
The screw connections on
the oxygen lines must be
tightened by hand only.
1. Open the valve of the oxygen cylinder
You can now read the pressure in the cylinder
from the gauge on the pressure reducer. For ex-
ample, a reading of 200 bar means that the cyl-
inder is full, whereas 100 bar means it is half full.
You should change the bottle in good time, e.g.
at below 50 bar, to ensure an adequate operat-
ing time.
2. Close the bottle valve again.
3. Watch the needle of the gauge on the pressure
reducer for approx. 1 minute. If it stays in the
same place, the system is free of leaks. If the
needle drops steadily, there is a leak somewhere.
Repairing leaks
1. Prepare a soap/water solution using non-
perfumed soap.
2. Wet all the screw and hose connections with the
solution. Bubbles will form at the site of the leak.
3. Depressurize the system:
To do this, first close the oxygen cylinder. Switch
on MEDUMAT Easy CPR briefly until the pres-
sure gauge on the O
switch off MEDUMAT Easy CPR again.
4. If leaks are discovered, the defective compo-
nents must be changed.
5. After changing, make a fresh check for leaks.
6. If it proves impossible to eliminate the leak, the
ventilator will have to be repaired.
cylinder reads "0". Then