A T
C P R
er to the electronic system if the main battery
fails. This makes it possible to trigger an alarm
even if the main battery fails.
The battery capacity is calculated to be sufficient for
power requirements under normal operating condi-
tions for the full period between the two-yearly ser-
vices. The batteries are completely replaced every
two years during servicing.
We recommend having the batteries changed only
by the manufacturer – WEINMANN Emergency – or
by qualified specialists expressly authorized by the
manufacturer. Special precautions need to be taken
during the change in order to protect the electronic
system.
In an emergency, proceed as follows:
Changing the main battery
1. Make sure the ventilator is switched off.
2. Open the battery compartment on the side of
the MEDUMAT Easy CPR (e.g. with a coin).
3. Remove the old 3.6 V lithium battery.
4. Put in a new battery. Make sure it is inserted the
right way round.
5. Close the battery compartment again.
Important
The 3.6 V lithium battery is a special battery for this
unit. Use only batteries supplied by
WEINMANN Emergency.
EN
Servicing
63