9.4 Change valve membrane in patient valve

Expiration
Spontaneous
arm
breathing
arm

9.5 Storage

EN
64
Servicing
If one of the valve membranes in the expiration or
spontaneous breathing arms of the patient valve is crin-
kled, sticky or misshapen, it must be changed.
Spontaneous breathing arm
1. Take the spontaneous breathing insert out of
the patient valve. To do so, push the two locking
lugs out of their seat, using a small screwdriver,
for example.
2. Pull the defective valve membrane out of the
spontaneous breathing insert using pointed
tweezers.
3. Put in a new valve membrane.
4. Push the spontaneous breathing insert back into
the patient valve.
Expiration tube
1. Use pointed tweezers to pull the defective valve
membrane out of the expiration arm.
2. Put in a new valve membrane.
If you do not intend to use MEDUMAT Easy CPR for
a long period, we recommend the following storage
precautions:
1. Clean and disinfect (see "6. Hygiene prepara-
tion" on page 42).
2. Store MEDUMAT Easy CPR in a dry place.
3. The battery can remain inside the unit even for
lengthy periods.