3.4 Attaching the nasal canula
Union nut
Connector
4. Hygienic preparation
Attention!
12
EN
Hygienic preparation
The nasal canula is made of transparent plastic. For
hygiene reasons it must be used for one patient only.
1. If necessary, screw the connector and union nut onto
the threaded connection of the OXYWAY Click.
2. Push the tube of the nasal canula onto the connnector
of the OXYWAY Click.
This product may contain disposable items. Dispos-
able items are intended to be used only once. So use these
items only once and do not reprocess them. Reprocessing
disposable items may impair the functionality and safety of
the product and lead to unforeseeable reactions as a result
of ageing, embrittlement, wear, thermal load, the effects
of chemical processes, etc.
If you need to clean the outside of the device, always use
a clean cloth. The cloth may be dry or moistened with
clean water.
In exceptional cases where wiping with disinfectant is es-
sential, take special care to ensure that no liquid enters the
OXYWAY Click. Otherwise there is a risk that residual dis-
infectant may enter the patient's airways and cause injury.
We recommend you to use the disinfectant TERRALIN
You are recommended to wear suitable gloves (e.g.
household or disposable gloves) during disinfection
procedures.
®
.