5. Functional check
Perform a functional check every 6 months. To do so,
proceed as follows:
1.
Make a visual check on the OXYWAY Click for
mechanical damage.
2.
Make sure that the dial head of the flowmeters is in
position "0".
3.
Connect the flowmeter to the central gas system.
4.
Check that the flowmeter is firmly engaged in the
socket of the central gas system.
5.
Now use the dial head to set all index positions in suc-
cession, and check at the oxygen outlet whether the
flow increases in line with the setting.
6.
Reset the click dial head to "0".
Important
OXYWAY Click is not a shutoff valve! Small amounts
of oxygen can still escape even when it is completely
closed.
7.
You can now start using the OXYWAY Click.
If the functional check reveals any faults, do not use the
OXYWAY Click.
Functional check
13