• It is recommended that the performance should be checked every 2 years and after mainte-
nance and repair, by retesting at least the requirements in limits of the error of the cuff pressure
indication and air leakage (testing at least at 50mmHg and 200mmHg).
• Please dispose of ACCESSORIES, detachable parts, and the ME EQUIPMENT according to
the local guidelines.
• Manufacturer will make available on request circuit diagrams, component part lists, descrip-
tions, calibration instructions,etc., to assist to service personnel in parts repair.
• The plug/adapter plug pins insulate the device from the main supply. Do not position the device
in a position where it is difficult to disconnect from the supply mains to safely terminate opera-
tion of ME equipment.
• The operator shall not touch output of batteries /adapter and the patient simultaneously.
• Cleaning :Dust environment may affect the performance of the unit. Please use the soft cloth to
clean the whole unit before and after use. Don't use any abrasive or volatile cleaners.
• The device doesn't need to be calibrated within two years of reliable service.
• If you have any problems with this device, such as setting up, maintaining or using, please
contact the SERVICE PERSONNEL of Gima. Don't open or repair the device by yourself in the
event of malfunctions. The device must only be serviced, repaired and opened by individuals at
authorized sales/service centers.
• Please report to Gima if any unexpected operation or events occur.
• Keep the unit out of reach of infants, young children or pets to avoid inhalation or swallowing of
small parts. It is dangerous or even fatal.
• Be careful to strangulation due to cables and hoses, particularly due to excessive length.
• At least 30 min required for ME equipment to warm from the minimum storage temperature
between uses until it is ready for intended use. At least 30 min required for ME equipment to
cool from the maximum storage temperature between uses until it is ready for intended use.
• This equipment needs to be installed and put into service in accordance with the information
provided in the ACCOMPANYING DOCUMENTS;
• Wireless communications equipment such as wireless home network devices, mobile phones,
cordless telephones and their base stations, walkie-talkies can affect this equipment and
should be kept at least a distance d away from the equipment. The distance d is calculated
by the MANUFACTURER from the 80MHz to 5.8 GHz column of Table 4 and Table 9 of IEC
60601-1-2:2014, as appropriate.
• Please use ACCESSORIES and detachable parts specified/ authorised by MANUFACTURE.
Otherwise, it may cause damage to the unit or danger to the user/patients.
• There is no luer lock connectors are used in the construction of tubing, there is a possibility that
they might be inadvertently connected to intravascular fluid systems, allowing air to be pumped
into a blood vessel.
• Please use the device under the environment which was provided in the user manual.
Otherwise, the performance and lifetime of the device will be impacted and reduced.
ENGLISH
34